Toshiba/Canon Image intensifier
The X-ray Image Intensifier E5877J-P1K is a high-performance industrial imaging solution designed to provide clear and detailed X-ray images. It features very high resolution and efficient soft X-ray penetration, making it ideal for non-destructive inspection applications in industries such as electronics, automotive, and aerospace. This lightweight and compact device enhances imaging efficiency and precision, ensuring reliable and accurate results.
Model list:
Toshiba E5877J-P1AK
Toshiba E5877J-P1K
Toshiba E5877-P1K
Canon E5877-P1K
Canon E5877J-P1K
Canon E5877J-P1AK
Key Features:
- Achieves very high resolution up to 110 Lp/cm in M mode for detailed imaging.
- Compact design with an overall length of 226 mm and a maximum diameter of 164 mm.
- Lightweight, weighing approximately 8.0 kg for easy integration and handling.
- CsI direct deposit input screen reduces X-ray scatter and improves detection efficiency.
- Anti-reflective coating on the output screen enhances image quality.
- Dual operation modes (N mode and M mode) for versatile imaging needs.
- Fast setup with a preheat time of less than 3 minutes.
- Compliance with international safety standards for reliable and safe operation.
Technical Datasheet:
Mechanical Specifications:
Mechanical Specifications:
- Overall Length: 226±3 mm
- Maximum Diameter: 164±3 mm
- Nominal Entrance Field Size: typ. 100 mm
- Output Image Diameter: 20±1 mm
- Weight: approx. 8.0 kg
Electrical Specifications:
- Anode Voltage: 25 to 30 kV (N mode), 25 to 30 kV (M mode)
- G3 Electrode Voltage: 1.5 to 3.5 kV (N mode), 9.0 to 12.0 kV (M mode)
- G2 Electrode Voltage: 350 to 750 V (N mode), 150 to 550 V (M mode)
- G1 Electrode Voltage: 50 to 250 V (N mode), 150 to 350 V (M mode)
- Photocathode Voltage: 0 V
Power Supply Specifications:
- DC Power Supply Input Voltage: 100 to 240 Vac 30VA
- DC Power Supply Output Voltage: 24±1 Vdc
- DC Power Supply Output Current: 400 mA max.
- Weight: approx. 1 kg
Environmental Specifications:
- Operating Temperature: 5°C to 40°C
- Storage Temperature: -15°C to 45°C
- Operating Humidity: 30% to 85% RH (non-condensing)
- Storage Humidity: 10% to 90% RH (non-condensing)
- Operating Pressure: 70 to 106 kPa
- Storage Pressure: 50 to 106 kPa
Performance Specifications:
- Central Resolution: 77 Lp/cm (N mode), 110 Lp/cm (M mode)
- Conversion Factor (Gx): 90 cd/m2 (typ.) for 10 μGy/s (N mode), 90 cd/m2 (typ.) for 1 mR/s (M mode)
- Contrast Ratio: 22:1 (10% area) (N mode)
Safety and Handling Specifications:
- Maximum Fluoroscopic X-ray Tube Voltage: 125 kVp
- Maximum Input Dose Rate: Continuous 4.37×10^-4 Gy/min [50 mR/min], Pulsed 8.73×10^-3 Gy/min [1 R/min]
- Minimum Distance Between Focal Spot and Entrance Plane: 73 cm